描述
CTS
Type: TSR-30/1000
Water-cooled
The type designation of the CTS cabinet series is structured as follows:
e.g., Type TSR-30/1000
1) T means: Temperature
2) S means: Stress Screening
3) R means: External Conditioning
4) -30 means: Lower temperature -30 °C
5) 1000 means: 1000 liters test chamber volume
Technical performance data:
A) TSR-30/1000 base unit without hose connections,
i.e., for use as a stand-alone temperature test chamber
Lhodpozgbntsfx Aifeh
Temperature tests:
Temperature range -40°C to +180°C
Final temperature -45°C
Temperature deviation ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5
Heating 6.0 K/min
Cooling 5.0 K/min
Heat radiation
approx. 5000 W at +20°C
approx. 4200 W at ±0°C
approx. 3000 W at -20°C
The technical data refer to an ambient temperature of +25 °C,
rated voltage 400V, without test specimen, without radiation, and without additional equipment.
Air volume flow rate
approx. 2000 m³/h
B) TSR-30/1000 base unit with hose connections,
for connection to an on-site external test chamber
Temperature tests:
Temperature range -30°C to +170°C
Temperature deviation ≤ ±0.1 to 0.5 K over time
Average temperature change rate,
measured with the circuit short-circuited
Heating 4.0 K/min
Cooling 4.0 K/min
in the range of +150°C to -10°C
The technical data refer to an ambient temperature of +25 °C,
rated voltage 400V, without test specimen, without radiation, and without additional equipment,
measured with the circuit short-circuited
Caution: The on-site external test chamber must be absolutely sealed and appropriately insulated;
any penetrations must be optimally sealed on-site.
Dimensions:
Test volume 1000 liters
Air circulation connection Ø 200 mm
Test chamber dimensions height 900 mm width 1000 mm depth 1100 mm
External dimensions height 1950 mm width 1270 mm depth 2245 mm
Technical design:
Refrigeration unit:
water-cooled, low-noise, hermetically sealed
stepless power adjustment by
electronic monitoring and control system,
environmentally friendly refrigerant R404A
Door:
fully opening door, door stop on the left,
one-hand operation, lockable.
Insert grid:
stainless steel, support area 970 x 980mm
Load per grid max. 35 kg
Total load with max. grid capacity 100 kg
Test chamber lighting: Halogen lamp 20 W, 12 V
Control:
- 32-bit controller
- LCD control panel
Operating data:
Electrical connection 400 V +6/-10 %, 3/N, 50 Hz
Rated power 13.3 kW
Protection class IP 22
Sound pressure level < 60 dB(A)
Weight approx. 850 kg
For your safety as a buyer, please note the following information:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See picture)
(Changes and errors in the technical data are subject to change!)
We will be happy to answer any further questions you may have by phone.
该广告为自动翻译,可能存在翻译错误。
Type: TSR-30/1000
Water-cooled
The type designation of the CTS cabinet series is structured as follows:
e.g., Type TSR-30/1000
1) T means: Temperature
2) S means: Stress Screening
3) R means: External Conditioning
4) -30 means: Lower temperature -30 °C
5) 1000 means: 1000 liters test chamber volume
Technical performance data:
A) TSR-30/1000 base unit without hose connections,
i.e., for use as a stand-alone temperature test chamber
Lhodpozgbntsfx Aifeh
Temperature tests:
Temperature range -40°C to +180°C
Final temperature -45°C
Temperature deviation ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5
Heating 6.0 K/min
Cooling 5.0 K/min
Heat radiation
approx. 5000 W at +20°C
approx. 4200 W at ±0°C
approx. 3000 W at -20°C
The technical data refer to an ambient temperature of +25 °C,
rated voltage 400V, without test specimen, without radiation, and without additional equipment.
Air volume flow rate
approx. 2000 m³/h
B) TSR-30/1000 base unit with hose connections,
for connection to an on-site external test chamber
Temperature tests:
Temperature range -30°C to +170°C
Temperature deviation ≤ ±0.1 to 0.5 K over time
Average temperature change rate,
measured with the circuit short-circuited
Heating 4.0 K/min
Cooling 4.0 K/min
in the range of +150°C to -10°C
The technical data refer to an ambient temperature of +25 °C,
rated voltage 400V, without test specimen, without radiation, and without additional equipment,
measured with the circuit short-circuited
Caution: The on-site external test chamber must be absolutely sealed and appropriately insulated;
any penetrations must be optimally sealed on-site.
Dimensions:
Test volume 1000 liters
Air circulation connection Ø 200 mm
Test chamber dimensions height 900 mm width 1000 mm depth 1100 mm
External dimensions height 1950 mm width 1270 mm depth 2245 mm
Technical design:
Refrigeration unit:
water-cooled, low-noise, hermetically sealed
stepless power adjustment by
electronic monitoring and control system,
environmentally friendly refrigerant R404A
Door:
fully opening door, door stop on the left,
one-hand operation, lockable.
Insert grid:
stainless steel, support area 970 x 980mm
Load per grid max. 35 kg
Total load with max. grid capacity 100 kg
Test chamber lighting: Halogen lamp 20 W, 12 V
Control:
- 32-bit controller
- LCD control panel
Operating data:
Electrical connection 400 V +6/-10 %, 3/N, 50 Hz
Rated power 13.3 kW
Protection class IP 22
Sound pressure level < 60 dB(A)
Weight approx. 850 kg
For your safety as a buyer, please note the following information:
The following points are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers are subjected to a documented test run.
Condition: used
Scope of delivery: (See picture)
(Changes and errors in the technical data are subject to change!)
We will be happy to answer any further questions you may have by phone.