带湿度的气候箱,100升Weiss
WKL 100/40 -40...+180°C
带湿度的气候箱,100升
Weiss
WKL 100/40 -40...+180°C
狀況
二手
位置
Borken 

顯示圖片
显示地图
机器数据
价格和位置
- 位置:
- Einsteinstraße 8a, 46325 Borken, Deutschland

撥打
优惠详情
- 產品ID:
- A20561415
- 參考編號:
- 24882
- 最後更新:
- 於 19.11.2025
描述
Weiss
Type: WKL 100/40
Climate test chamber, air-cooled
with humidity control, -40 to +180°C
Specifications for temperature tests:
Temperature range: -40 to +180°C
Temperature uniformity after stabilization:
Temporal temperature fluctuation in the center of the usable space: ±0.3 to ±1 K
Spatial temperature deviation: ±0.5 to ±2 K
Temperature change rate:
Heating: 3.5 K/min
Cooling: 5 K/min
Heat compensation: max. 1,100 W
Specifications for climate (humidity) tests:
Humidification system data
Tank capacity: approx. 13 l
Water quality: Demineralized
pH value: 6 – 7
Conductivity: max. 20 µS/cm
Water consumption at constant +40°C, 92% r.h.: 2 l / 24 h
Temperature range: +10 to +95°C
Humidity range: 10 to 98% r.h.
Temperature and humidity deviations (after stabilization):
Temporal temperature fluctuation in the chamber center: ±0.3 to ±0.5 K
Spatial temperature deviation: ±0.5 to ±1.5 K
Temporal humidity fluctuation in the chamber center: ±1 to ±3% r.h.
Mechanical load data:
Maximum floor load in test chamber: 25 kg
Per shelf: 10 kg
Total rack load: 50 kg
Right-side port: 50 mm ∅
Left-side port: 50 mm ∅
Interfaces: Ethernet (RJ45), USB
Operating data:
Chamber lighting: Halogen bulb, 24V, 50W
Rated voltage: 1/N / PE AC 230 V ±10% 50 Hz
Rated power: 3.0 kW
Rated current: 13.0 A
Site-side fuse: 16A slow-blow
Protection class:
Control cabinet IP 22, control panel IP 54
Heat output, air-cooled chambers:
Maximum heat output to environment: 2,000 W
Noise measurement:
Sound pressure level (free field measurement, 1 m distance in front, 1.6 m height):
approx. 56 dB(A)
Kedpfxsxxtuij Agfsnd
Test chamber volume: approx. 100 liters
Weight: 190 kg
Refrigerant: R-452a
For your safety as a buyer, please note the following information:
The following actions are performed on all chambers we offer prior to sale:
1. Functional testing and replacement of necessary components
2. If necessary, refilling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data and descriptions reserved!)
We are happy to answer further questions by phone.
廣告是自動翻譯的,翻譯中可能會出現一些錯誤。
Type: WKL 100/40
Climate test chamber, air-cooled
with humidity control, -40 to +180°C
Specifications for temperature tests:
Temperature range: -40 to +180°C
Temperature uniformity after stabilization:
Temporal temperature fluctuation in the center of the usable space: ±0.3 to ±1 K
Spatial temperature deviation: ±0.5 to ±2 K
Temperature change rate:
Heating: 3.5 K/min
Cooling: 5 K/min
Heat compensation: max. 1,100 W
Specifications for climate (humidity) tests:
Humidification system data
Tank capacity: approx. 13 l
Water quality: Demineralized
pH value: 6 – 7
Conductivity: max. 20 µS/cm
Water consumption at constant +40°C, 92% r.h.: 2 l / 24 h
Temperature range: +10 to +95°C
Humidity range: 10 to 98% r.h.
Temperature and humidity deviations (after stabilization):
Temporal temperature fluctuation in the chamber center: ±0.3 to ±0.5 K
Spatial temperature deviation: ±0.5 to ±1.5 K
Temporal humidity fluctuation in the chamber center: ±1 to ±3% r.h.
Mechanical load data:
Maximum floor load in test chamber: 25 kg
Per shelf: 10 kg
Total rack load: 50 kg
Right-side port: 50 mm ∅
Left-side port: 50 mm ∅
Interfaces: Ethernet (RJ45), USB
Operating data:
Chamber lighting: Halogen bulb, 24V, 50W
Rated voltage: 1/N / PE AC 230 V ±10% 50 Hz
Rated power: 3.0 kW
Rated current: 13.0 A
Site-side fuse: 16A slow-blow
Protection class:
Control cabinet IP 22, control panel IP 54
Heat output, air-cooled chambers:
Maximum heat output to environment: 2,000 W
Noise measurement:
Sound pressure level (free field measurement, 1 m distance in front, 1.6 m height):
approx. 56 dB(A)
Kedpfxsxxtuij Agfsnd
Test chamber volume: approx. 100 liters
Weight: 190 kg
Refrigerant: R-452a
For your safety as a buyer, please note the following information:
The following actions are performed on all chambers we offer prior to sale:
1. Functional testing and replacement of necessary components
2. If necessary, refilling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data and descriptions reserved!)
We are happy to answer further questions by phone.
廣告是自動翻譯的,翻譯中可能會出現一些錯誤。